4- 122
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
TSQ
The TSQ command returns the time data from when the TSR command is sent until
this command is sent. The time data will be put in the data output buffer as same as
the measurement data.
This command is effective for all measurement modes, regardless of the TSC
setting.
This command is effective for ASCII data output format. Refer to
.
Syntax
TSQ
Example
Statements
OUTPUT @E5270;"TSQ"
ENTER @E5270 USING "#,5X,13D,X";Time
PRINT "Time=";Time;"s"
TSR
The TSR command clears the timer count.
This command is effective for all measurement modes, regardless of the TSC
setting.
This command is effective for ASCII data output format. Refer to
.
Syntax
TSR
Remarks
To read the time data with the best resolution (100
μ
s), the timer must be cleared
every 100 sec or less for the FMT1, 2, or 5 data output format, or
every 1000 sec or less for the FMT 11, 12, 15, 21, 22, or 25 data output format.
Example
Statements
OUTPUT @E5270;"TSR"