4- 62
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
IN
The IN command sets the specified channel to 0 V with an output range change.
Syntax
IN [
chnum
[,
chnum
...[,
chnum
]...]]
A maximum of eight channels can be set.
Parameters
chnum
:
Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
.
If you specify multiple
chnum
s, the E5260/E5270 sets the channels to 0 V in the
specified order.
If you do not specify
chnum
, this command sets all channels to 0 V in order from
higher voltage range (output or measurement range) to lower voltage range.
Remarks
The IN command sets the specified channels to the following conditions, which are
the same as the conditions after executing the CN command.
Example
Statements
OUTPUT @E5270;"IN"
OUTPUT @E5270;"IN 1,2,3,5,6"
Item
SMU
GNDU
Source Mode
V
Output Voltage
0 V
0 V
V Range
20 V
I Compliance
100
μ
A
I Range
100
μ
A
Filter
Not changed
Power Consumption
0 W