WNU?
The WNU? query command returns the number of sweep steps specified by the
sweep command (WI, WV, PWI or PWV), and stores the results in the output data
buffer (query buffer).
Execution
Conditions
If you want to know the number of steps for a pulsed sweep, you must execute an
“MM 4” command before using this command, otherwise the number of steps for
the staircase sweep is reported.
Syntax
WNU?
Query Response
number of sweep steps
<CR/LF^EOI>
Example
Statement
OUTPUT @E5270;"WNU?"