3- 40
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Examples
Linear Search Measurements
Linear Search Measurements
To perform linear search measurements, use the following commands.
The LSV and LSI commands clear the previous source settings.
Send the LSI command before sending the LSSI command.
Send the LSV command before sending the LSSV command.
The LSI/LSSV commands or LSV/LSSI commands cannot be used together.
Function
Command
Parameters
Enables channels
CN
[
chnum
... [,
chnum
] ... ]
Disables channels
CL
[
chnum
... [,
chnum
] ... ]
Sets filter ON/OFF
[FL]
mode
[,
chnum
... [,
chnum
] ... ]
Sets series resistor ON/OFF
[SSR]
chnum
[,
mode
]
Sets integration time
(Keysight E5270B can use
AAD/AIT instead of AV.)
[AV]
number
[,
mode
]
[AAD]
chnum
[,
type
]
[AIT]
type
,
mode
[,
N
]
Sets measurement mode
MM
14
Selects output data
[LSVM]
output_data
Sets timing parameters
[LSTM]
hold
,
delay
Sets auto abort function
[LSM]
abort
[,
post
]
Sets current search or
voltage search condition
LGI or
LGV
chnum
,
mode
,
range
,
target
Sets voltage source or
current source
LSV or
LSI
chnum
,
range
,
start
,
stop
,
step
[,
comp
]
Sets synchronous voltage
source or current source
[LSSV] or
[LSSI]
chnum
,
polarity
,
offset
[,
comp
]
Forces constant voltage
DV, TDV
chnum
,
range
,
output
[,
comp
[,
polarity
[,
crange
]]]
Forces constant current
DI, TDI
Executes measurement
XE
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......