1- 34
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Basics
Data Output Format
D
Value of
Data
,
Source_data
,
Data_search
, or
Data_sense
.
This value is expressed in 17-bit binary data. The value can be calculated by the
following formula.
Measurement data =
Count
×
Range
/50000
Source data =
Count
×
Range
/20000
where,
Count
is the decimal value of
D
, and
Range
is the measurement range or
output range indicated by
C
.
For the current data, the measurement range or output range value can be calculated
by the following formula:
Current measurement/output range (A) = 10^(
C
-20)
If the top bit of the binary data is 0,
Count
is positive and equal to the decimal value
of the 16-bit binary data that follows the top bit.
If the top bit is 1,
Count
is negative. Calculate
Count
by subtracting 65536
(10000000000000000 in binary) from the decimal value of the 16-bit binary data.
Example:
If the output binary data is:
11010110000100111000100000000001
then,
Data type:
Current measurement data (
A
=1,
B
=1)
Range
:
1 nA=1E–9 A (
C
=01011)
Count
:
5000 (
D
=00001001110001000)
Status:
Normal condition (
E
=000)
Channel:
SMU1 (channel number 1) (
F
=00001)
Measurement data
= 5000
×
1E–9/5E+4 = 100 pA
NOTE
B
=1 and
C
=10100 means that HPSMU used 1 A range or MPSMU used 200 mA
range. Then use
Range
=1 to calculate the data for both HPSMU and MPSMU.
Range
=0.2 is not available even if the range value is 200 mA.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......