3- 70
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Examples
Reading Binary Output Data
Table 3-18
High-Speed Spot Measurement Example to read binary data
Sub perform_meas(ByVal session As IMessage, ByVal t() As Integer) ’1
Dim i As Integer = 0 ’t(0): Drain
Dim j As Integer = 0 ’t(1): Gate
Dim nop1 As Integer = 1 ’t(2): Source
Dim nop2 As Integer = 1 ’t(3): Substrate
Dim data(nop2 - 1, nop1 - 1) As String
Dim value As String = "Id (mA), Status"
Dim fname As String = "C:\Keysight\prog_ex\data16.txt"
Dim title As String = "Measurement Result"
Dim msg As String = "No error."
Dim err As Integer = 0
Dim vd As Double = 3 ’13
Dim vg As Double = 1
Dim idcomp As Double = 0.05
Dim igcomp As Double = 0.01
Dim orng As Integer = 0
Dim mrng As Integer = 0
session.WriteString("FMT 3" & vbLf)
session.WriteString("AV 10,1" & vbLf) ’sets number of samples for 1 data
session.WriteString("FL 0" & vbLf) ’sets filter off
session.WriteString("DV " & t(3) & ",0,0,0.1" & vbLf) ’out= 0 V, comp= 0.1 A
session.WriteString("DV " & t(2) & ",0,0,0.1" & vbLf) ’out= 0 V, comp= 0.1 A
session.WriteString("DV " & t(1) & "," & orng & "," & vg & "," & igcomp & vbLf)
session.WriteString("DV " & t(0) & "," & orng & "," & vd & "," & idcomp & vbLf)
session.WriteString("ERR? 1" & vbLf) : err = session.ReadString(4 + 2) ’28
If err <> 0 Then session.WriteString("DZ" & vbLf) : GoTo Check_err
session.WriteString("TI " & t(0) & "," & mrng & vbLf)
Dim dat() As Byte = session.Read(4 + 2) ’4 byte data + terminator ’31
Dim status As Integer = dat(3) And 224 : status = status / 32 ’224=128+64+32
If status <> 0 Then session.WriteString("DZ" & vbLf) : GoTo Check_err
Dim type As Integer = dat(0) And 128 : type = type / 128 ’0:source, 1:meas
Dim mode As Integer = dat(0) And 64 : mode = mode / 64 ’0:voltage, 1:current
Dim sign As Integer = dat(0) And 1 ’0:positive, 1:negative
Dim rng As Integer = dat(0) And 62 : rng = rng / 2 ’62=32+16+8+4+2
Dim count As Integer = dat(1) * 256 + dat(2)
Dim chan As Integer = dat(3) And 31 ’31=16+8+4+2+1
If sign = 1 Then count = count - 65536 ’65536 = 10000000000000000 (17 bits)
Line
Description
2 to 11
Declares variables used through the project. And sets the proper values.
13 to 18
Declares variables and sets the value.
20 to 22
Sets the data output format and A/D converter. Also sets the SMU filter off.
23 to 26
Applies voltage to device.
28 to 29
Checks if an error occurred. If an error is detected, forces 0 V and goes to Check_err.
30 to 31
Performs the high-speed spot measurement. And stores the returned binary data (four
bytes) into the
dat
array variable.
33 to 41
Picks up the elements, status, type, mode, sign, rng, count, and chan, included in the
returned binary data.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......