![Keysight Technologies E5260 Series Programming Manual Download Page 316](http://html1.mh-extra.com/html/keysight-technologies/e5260-series/e5260-series_programming-manual_1961978316.webp)
4- 132
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
WI
The WI command specifies the staircase sweep current source and its parameters.
This command also clears the WV, WSV, WSI, and WNX command settings.
This command setting is cleared by the WV command.
Syntax
•
For Staircase Sweep Measurement:
WI
chnum
,
mode
,
range
,
start
,
stop
,
step
[,
Vcomp
[,
Pcomp
]]
•
For Staircase Sweep with Pulsed Bias Measurement:
WI
chnum
,
mode
,
range
,
start
,
stop
,
step
[,
Vcomp
]
Parameters
chnum
:
Sweep source channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
.
mode
:
Sweep mode. Integer expression. Only linear sweep (
mode
=1 or 3) is
available for the staircase sweep with pulsed bias.
1: Linear sweep (single stair, start to stop.)
2: Log sweep (single stair, start to stop.)
3: Linear sweep (double stair, start to stop to start.)
4: Log sweep (double stair, start to stop to start.)
range
:
Ranging type for staircase sweep current output. Integer expression.
See
For the linear sweep, the E5260/E5270 uses the minimum range that
covers both
start
and
stop
values to force the staircase sweep current.
For the log sweep, the E5260/E5270 uses the minimum range that
covers the output value, and changes the output range dynamically.
For the limited auto ranging, the instrument never uses the range less
than the specified range.
start
,
stop
:
Start or stop current (in A). Numeric expression. See
.
start
and
stop
must have the same polarity for
log
sweep.
0 to ±0.1, 0 to ±0.2 for E5291A, or 0 to ±1 for E5280B/E5290A
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......