Keysight E5260/E5270 Programming Guide, Edition 4
4- 75
Command Reference
Command Reference
LSSI
The LSSI command specifies and sets the synchronous current source for the linear
search measurement. The synchronous source output will be:
Synchronous source output =
polarity
×
LSI source
offset
where the LSI source output is the output set by the LSI command.
This command setting is cleared by the LSV/LSI command.
Execution
Conditions
The LSI command must be entered before this command.
Syntax
LSSI
chnum
,
polarity
,
offset
[,
Vcomp
]
Parameters
chnum
:
Synchronous source channel number. The value must be slot number
where the module has been installed. Integer expression. 1 to 8 (1 and 5
are not available for HPSMU). See
.
polarity
:
Polarity of the LSSI output for the LSI output.
0 (negative): LSSI output =
−
LSI
offset
1 (positive): LSSI output = LSI
offset
offset
:
Offset current (in A). Numeric expression.
0 to ±0.1, 0 to ±0.2 for E5291A, or 0 to ±1 for E5280B/E5290A
Both primary and synchronous search sources will use the same output
range. So check the output range set to the LSI command to determine
the synchronous source outputs.
Vcomp
:
Voltage compliance value (in V). Numeric expression. If you do not
specify
Vcomp,
the previous value is set.
Example
Statements
OUTPUT @E5270;"LSSI 1,1,1E-6,5"
See Also
for the source output value, output range, and the
available compliance values.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......