Keysight E5260/E5270 Programming Guide, Edition 4
4- 65
Command Reference
Command Reference
LGV
The LGV command specifies the voltage monitor channel and its search parameters
for the linear search measurement.
This command ignores the RV command setting.
This command setting is cleared by the LGI command.
Syntax
LGV
chnum
,
mode
,
range
,
target
Parameters
chnum
:
Search monitor channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
.
mode
:
Search mode. Integer expression.
0 : If the measured value
≤
target
, it is the search result data.
1 : If the measured value
≥
target
, it is the search result data.
range
:
Measurement ranging type. Integer expression. The measurement range
will be set to the minimum range that covers the
target
value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
target
:
Search target voltage (in V). Numeric expression.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Example
Statements
OUTPUT @E5270;"LGV 1,2,12,3"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......