4- 82
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
MM
The MM command specifies the measurement mode and the channels used for
measurements. This command must be entered to specify the measurement mode.
For the high speed spot measurements, do not enter the MM command.
Syntax
•
For spot, staircase sweep, and multi channel sweep:
MM
mode
,
chnum
[,
chnum
[,
chnum
...[,
chnum
]...]]
A maximum of eight channels can be set.
•
For pulsed spot, pulsed sweep, and staircase sweep with pulsed bias:
MM
mode
,
chnum
•
For binary search and linear search:
MM
mode
•
For quasi pulsed spot:
MM
mode
[,
chnum
]
Parameters
mode
:
Measurement mode. Integer expression. 1 to 5, 9, and 14 to 16.
chnum
:
Measurement channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
.
mode
Measurement mode
Related source setup command
1
Spot
DI, DV
2
Staircase sweep
WI, WV, WT, WM, WSI, WSV
3
Pulsed spot
PI, PV, PT
4
Pulsed sweep
PWI, PWV, PT, WM, WSI, WSV
5
Staircase sweep with
pulsed bias
WI, WV, WM, WSI, WSV, PI, PV, PT
9
Quasi-pulsed spot
BDV, BDT, BDM
14
Linear search
LSV, LSI, LGV, LGI, LSM, LSTM,
LSSV, LSSI, LSVM
15
Binary search
BSV, BSI, BGV, BGI, BSM, BST,
BSSV, BSSI, BSVM
16
Multi channel sweep
WI, WV, WT, WM, WNX
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......