Command Reference
Command Reference
Figure 4-1
Binary Search Source Output Control Mode
Normal mode
The operation of the normal mode is explained below:
1. The source channel forces the Start value, and the monitor channel executes a
measurement.
2. The source channel forces the Stop value, and the monitor channel executes a
measurement.
If the search target value is out of the range between the measured value at the
Start value and the measured value at the Stop value, the search stops.
3. The source channel forces the Stop
-
D/2 value (or Stop+D/2 if Start>Stop), and
the monitor channel executes a measurement.
If the search stop condition is not satisfied, the measured data is used to decide
the direction (+ or –) of the next output change. The value of the change is
always half of the previous change.
4. Repeats the output change and measurement until the search stop condition is
satisfied.
For information on the search stop condition, see
. If the output
change value is less than the setting resolution, the search stops.
Start
Stop
D = | Stop - Start |
+D/4
-D/2
-D/8
-D/16
(1) Normal output
Search stopped
+D/32
+D/64
: Measurement
Start
Stop
+D/4
+D/2
-D/8
-D/16
(2) Cautious output
Search stopped
+D/32
+D/64
Time
Voltage or current
Hold time
Delay time
Hold time
Delay time
Delay time
Delay time
Trigger
Trigger
BSM, BST, and
BSV or BSI
BSM, BST, and
BSV or BSI
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......