Keysight E5260/E5270 Programming Guide, Edition 4
4- 49
Command Reference
Command Reference
DIAG?
The DIAG? command starts the diagnostics, and returns the results in ASCII format.
Before starting the diagnostics, refer to Remarks below.
After the DIAG? command, read the results soon.
Syntax
DIAG?
item
[,
pause
]
Parameters
item
:
Diagnostics item. Integer expression. 1 to 5.
pause
:
Pauses before starting diagnostics or not. Integer expression. 0 or 1.
This parameter is effective for
item
=1 and 4.
0: Keysight E5260/E5270 starts diagnostics immediately.
1: Keysight E5260/E5270 starts diagnostics when the
Enter
key is
pressed.
If you do not specify
pause
, 1 is set.
For
pause
=1, you can abort execution of the diagnostics by pressing the
Exit
key while a message is being displayed on the LCD.
Query Response
result
<CR/LF^EOI>
0: Passed.
1: Failed.
2: Aborted.
Example
Statements
OUTPUT @E5270;"DIAG? 1,1"
ENTER @E5270;A
item
Description
pause
1
Trigger In/Out diagnostics.
effective
2
Front panel key diagnostics.
n.a
3
High voltage LED diagnostics.
n.a
4
Digital I/O diagnostics.
effective
5
Beeper diagnostics.
n.a
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......