Keysight E5260/E5270 Programming Guide, Edition 4
2- 3
Remote Mode Functions
Measurement Modes
Measurement Modes
The Keysight E5260/E5270 provides the following measurement modes.
•
•
•
“Staircase Sweep Measurements”
•
“Multi Channel Sweep Measurements”
•
•
“Staircase Sweep with Pulsed Bias Measurements”
•
“Quasi-Pulsed Spot Measurements”
•
•
NOTE
About Search Measurements
The E5260/E5270 supports search measurement to find a point on an I-V curve
where a specified condition is satisfied. For example, it searches for a breakdown
voltage or threshold voltage at a specified current.
Search measurements are performed by one or two SMUs. For two SMUs, one is the
search channel, and the other is a sense channel. When one SMU is used, it serves as
both search and sense channel.
Basically, the search channel forces voltage or current until the search stop condition
is satisfied.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......