Keysight E5260/E5270 Programming Guide, Edition 4
4- 21
Command Reference
Command Reference
AIT
This command is available for the Keysight E5270B, and is used to set the
integration time or the number of averaging samples of the A/D converter (ADC)
for each ADC type.
This command setting is ignored by the pulsed spot, pulsed sweep, and staircase
sweep with pulsed bias measurements.
Execution
Conditions
Enter the AAD command to specify the ADC type for each measurement channel.
This command is not available for the Keysight E5260 series.
Syntax
AIT
type
,
mode
[,
N
]
Parameters
type
:
A/D converter type. Integer expression.
0: High-speed A/D converter.
1: High-resolution A/D converter.
mode
:
ADC operation mode. Integer expression. Initial value is 0.
0: Auto mode.
1: Manual mode.
2: Power line cycle (PLC) mode.
N
:
Coefficient used to define the integration time or the number of
averaging samples. Integer expression. See
.
Example
Statements
OUTPUT @E5270;"AIT 0,2,1"
OUTPUT @E5270;"AIT 1,1,10"