4- 130
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
WAT
This command sets the source wait time and the measurement wait time as shown in
. The wait time is given by the following formula:
wait time
=
N
×
initial wait time
+
offset
where
initial wait time
is the time the Keysight E5260/E5270 initially sets and you
cannot change. The
initial source wait time
is not same as the
initial measurement
wait time
. The wait time settings are effective for all modules.
Syntax
WAT
type
,
N
[,
offset
]
Parameters
type
Type of the wait time. Integer expression. 1 or 2.
1: Source wait time (before changing the output value).
2: Measurement wait time (before starting the measurement).
N
Coefficient for
initial wait time
. Numeric expression.
0 to 10, resolution 0.1. Initial value is 1.
offset
Offset for the wait time. Numeric expression.
0 to 1 sec, resolution 0.0001. Default value is 0.
Example
Statements
OUTPUT @E5270;"WAT 1,.7"
OUTPUT @E5270;"WAT 2,0,.01"
Figure 4-4
Source/Measurement Wait Time
NOTE
The wait time can be ignored if it is shorter than the delay time.
Measurement wait time
Time
: Measurement
Hold time
Delay time
Source wait time
Step delay time
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......