4- 60
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
Table 4-11
FMT format parameter
12 digits data will be
sn.nnnnnEsnn
,
snn.nnnnEsnn
, or
snnn.nnnEsnn
.
13 digits data will be
sn.nnnnnnEsnn
,
snn.nnnnnEsnn
, or
snnn.nnnnEsnn
.
where,
s
is
+
or
-
,
E
is exponent symbol, and
n
means one digit number.
NOTE
For binary data output format, the time stamp function is not available. Refer to
“Data Output Format” on page 1-22
.
format
Data format
Terminator
1
a
a. Compatible with the Keysight 4142B data output format.
ASCII (12 digits data with header)
<CR/LF^EOI>
2
a
ASCII (12 digits data without header)
<CR/LF^EOI>
3
a
binary
<CR/LF^EOI>
4
a
binary
<^EOI>
5
a
ASCII (12 digits data with header)
,
11
ASCII (13 digits data with header)
<CR/LF^EOI>
12
ASCII (13 digits data without header)
b
<CR/LF^EOI>
15
ASCII (13 digits data with header)
,
21
ASCII (13 digits data with header)
b
b. Compatible with the Keysight 4155/4156 FLEX mode ASCII data.
<CR/LF^EOI>
22
ASCII (13 digits data without header)
b
<CR/LF^EOI>
25
ASCII (13 digits data with header)
b
,
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......