Keysight E5260/E5270 Programming Guide, Edition 4
4- 45
Command Reference
Command Reference
CMM
The CMM command sets the SMU measurement operation mode. This command is
not available for the high speed spot measurement.
Syntax
CMM
chnum
,
mode
Parameters
chnum
:
Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
.
mode
:
SMU measurement operation mode. Integer expression.
•
0: Compliance side measurement (initial setting).
•
1: SMU always performs current measurement.
•
2: SMU always performs voltage measurement.
•
3: Force side measurement.
If
mode
=0, SMU measures current when it forces voltage, or measures voltage when
it forces current.
If
mode
=3, SMU measures current when it forces current, or measures voltage when
it forces voltage.
The
mode
setting is kept until the
mode
is changed by this command. If you want to
return it to the initial setting, enter the CMM command with
mode
=0.
Example
Statements
OUTPUT @E5270;"CMM 1,1"