Keysight E5260/E5270 Programming Guide, Edition 4
4- 127
Command Reference
Command Reference
TV
The TV command performs the high speed spot measurement, and returns the
measurement data. The command starts a voltage measurement regardless of the
SMU operation mode, trigger mode (TM command), and measurement mode (MM
command).
Execution
Conditions
CN command has been executed for the specified channel.
Syntax
TV
chnum
[,
range
]
Parameters
chnum
:
Measurement channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
.
range
:
Measurement range or ranging type. Integer expression. If you select
the fixed range, the instrument performs measurement by using the
specified range. For the auto or limited auto ranging, the measurement
range will be set to the minimum range that covers the measured
values. However, the instrument never uses the range less than the
specified range for the limited auto ranging. See
.
If you do not specify the
range
parameter for current output channels,
the channel uses the minimum range that covers the compliance value.
The
range
parameter is meaningless for voltage output channels. The
measurement ranging type is always same as the output ranging type.
Example
Statements
OUTPUT @E5270;"TV 1"
ENTER @E5270 USING "#,3X,12D,X";Vdata
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......