4- 138
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
Remarks
To set multiple sweep sources, enter the WI or WV command at first, and enter the
WNX command once or more. Then the
N
value and the
chnum
value must be
unique for each WNX command. If you set the value used to the previous command,
the previous command setting is cleared, and the last command setting is effective.
Sweep sources simultaneously start output by a trigger such as the XE command.
However, if a sweep source sets power compliance or forces logarithmic sweep
current, the sweep sources start output in the order specified by the
N
value. Then
the first output is forced by the channel set by the WI or WV command.
If you use multiple measurement channels, the channels that use the high-speed A/D
converter with the fixed ranging mode start measurement simultaneously, then other
channels start measurement in the order defined in the MM command.
Example
Statements
OUTPUT @E5270;"WNX 2,3,1,12,0,3,1E-3,2E-3"
OUTPUT @E5270;"WNX 3,4,2,0,1E-3,1E-2,3"