1- 10
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Basics
Getting Started
To Set the Integration Time
To adjust the balance of the measurement accuracy and the measurement speed,
change the integration time or the number of averaging samples of the A/D
converter (ADC) by using the AV command. The AV command is compatible with
the AV command of the Keysight 4142B.
For accurate and reliable measurement, set the integration time longer or set the
number of samples larger. For details about the integration time settings, refer to
Chapter 4, “Command Reference.”
The Keysight E5270B has the following two types of the A/D converter. Use the
AAD command to select the type of ADC, and use the AIT command to set the
integration time or the number of samples.
Example
The following example sets the number of samples to 10 for the A/D converter
(high-speed ADC for the Keysight E5270B).
OUTPUT @E5270;"AV 10,1"
The following example is for the Keysight E5270B, and sets the power line cycle
mode (PLC) for both the high-speed ADC and the high-resolution ADC. And
channel 1 uses the high-resolution ADC and other channels use the high-speed
ADC.
OUTPUT @E5270;"*RST"
OUTPUT @E5270;"AIT 0,2,1"
OUTPUT @E5270;"AIT 1,2"
OUTPUT @E5270;"AAD 1,1"
Type of ADC
Description
High-speed ADC
Effective for the high speed measurement. In the multi
channel sweep measurement mode (MM16), multiple
measurement channels can perform synchronous
measurements.
The number of averaging samples must be set by the
AIT or AV command.
High-resolution ADC
Effective for the accurate measurement. Cannot be used
for the pulsed measurement channel and the
simultaneous measurement channel.
The integration time must be set by the AIT command.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......