Keysight E5260/E5270 Programming Guide, Edition 4
4- 139
Command Reference
Command Reference
WS
The WS command causes the E5260/E5270 to enter a wait state until the
E5260/E5270 receives an external trigger from the Ext Trig In terminal. To set the
trigger logic (initial value: negative), send the TGP command for the Ext Trig In
terminal.
To end a wait state before the trigger, execute the AB or *RST command.
Syntax
WS [
mode
]
Parameters
mode
:
Waiting mode. Integer expression. 1 or 2.
If this parameter is not specified,
mode
is set to 1.
Remarks
The E5260/E5270 checks its trigger flag to confirm the present trigger status,
received or none. To clear the trigger flag:
•
Enter the *RST or device clear command (HP BASIC CLEAR statement).
•
Enter the TM3 command.
•
Enter the TM1, TM2, or TM4 command to change the mode from TM3.
•
Enter the OS command.
•
Trigger the E5260/E5270 to start measurement via the Ext Trig In terminal.
•
Trigger the E5260/E5270 to recover from wait state set by the WS command via
the Ext Trig In terminal.
Example
Statements
OUTPUT @E5270;"WS 2"
mode
Description
1
Continues the operation if an external trigger was
already received. Otherwise, the E5260/E5270
immediately goes into a wait state for an external
trigger.
2
In any condition, the E5260/E5270 immediately goes
into a wait state for an external trigger.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......