Keysight E5260/E5270 Programming Guide, Edition 4
2- 13
Remote Mode Functions
Measurement Modes
7. After the sweep measurement, the pulsed source forces the pulse base value, and
the sweep source forces the start or stop value, as specified by the WM
command, and keeps it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE
The WM command sets the automatic abort function and the output after
measurement. The WV command sets the sweep voltage, and the WI command sets
the sweep current. The start and stop values must have the same polarity for log
sweep.
The PT command sets the pulse timing parameters, such as pulse width and pulse
period. The PV command sets the voltage pulse, and the PI command sets current
pulse.
To Use
Synchronous
Sweep Source
One more channel can be set up as a sweep source that has the output synchronized
with the staircase sweep. Refer to
“Synchronous Output” on page 2-21
.
After the measurement, the synchronous sweep source forces the start or stop value,
as specified by the WM command, and keeps it.
Figure 2-9
Synchronous Sweep
Voltage or current
Start value
Time
Stop value
Primary sweep
Synchronous sweep
Previous value
Trigger (e.g. XE)
WT/WM/WV/WI
WSV/WSI
Voltage or current
Start value
Stop value
Previous value
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......