Keysight E5260/E5270 Programming Guide, Edition 4
2- 9
Remote Mode Functions
Measurement Modes
5. After measurement, the sweep source waits for the rest of the step delay time if
it is set, and the sweep source changes the output value.
6. The E5260/E5270 repeats 4 and 5 for all sweep steps.
7. After the sweep measurement, the sweep sources force the start or stop value, as
specified by the WM command, and keep it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE
The WT command sets the hold time, delay time, and step delay time. The WM
command sets the automatic abort function and the output after measurement. The
WV/WI command sets the output of the first sweep source, and the WNX command
sets the output of the
n
th (
n
=2 to 8) sweep source. The start and stop values must
have the same polarity for log sweep.
To Stop Sweep
Output
An automatic abort function is available. Refer to
Even if the automatic abort function is disabled, the E5260/E5270 automatically
stops measurement if power compliance is enabled for the sweep source and the
power compliance or an automatic abort condition is detected.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......