4- 116
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
TGPC
The TGPC command clears the trigger setting of the specified ports.
Syntax
TGPC [
port
[,
port
...[,
port
]...]]
A maximum of 18 ports can be set. If no port is specified, the TGPC command
clears the setting of the all ports; Ext Trig In, Ext Trig Out, and digital I/O ports 1 to
16.
Parameters
port
:
Trigger port number. Integer expression.
-
1,
-
2, or 1 to 16.
-
1: Ext Trig In terminal.
-
2: Ext Trig Out terminal.
1 to 16: Port 1 to 16 of the digital I/O terminal.
Remarks
The TGPC command sets the trigger ports as shown below.
Ext Trig In
Same as after TGP -1,1,2,0 command execution.
Ext Trig Out
Same as after TGP -2,2,2,0 command execution.
Digital I/O Ports
No trigger function is available. The ERS? and ERC commands
are available for the port control.
This is not same as the condition set by the *RST command that sets the ports as
shown below.
Ext Trig In
Same as after TGP -1,1,2,1 command execution.
Ext Trig Out
Same as after TGP -2,2,2,1 command execution.
Digital I/O Ports
No trigger function is available. The ERS? and ERC commands
are available for the port control.
Example
Statements
OUTPUT @E5270;"TGPC -1,-2,1,2"
See Also
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......