TGP
The TGP command enables the trigger function for the terminal specified by the
port
parameter. For the trigger function, refer to
“Trigger Function” on page 2-30
.
Syntax
TGP
port
,
terminal
,
polarity
[,
type
]
Parameters
port
:
Trigger port number. Integer expression.
-
1,
-
2, or 1 to 16.
-
1: Ext Trig In terminal.
-
2: Ext Trig Out terminal.
1 to 16: Port 1 to 16 of the digital I/O terminal.
terminal
:
Terminal type. Integer expression. 1 or 2.
1: Trigger input. Not available for
port
=
-
2.
2: Trigger output. Not available for
port
=
-
1.
polarity
:
Trigger logic. Integer expression. 1 or 2.
1: Positive logic.
2: Negative logic.
type
:
Trigger type. Integer expression. 0, 1, 2, or 3. Selects the function of the
trigger port. See
If this parameter is not specified,
type
is set to 0.
Remarks
The function of type=0 is effective for all trigger ports regardless of the
type
value.
Then the PA and WS commands are used for the Ext Trig In terminal, and the OS
command is used for the Ext Trig Out terminal. Also the PAX and WSX commands
are used for the trigger input ports set by the TGP command, and the OSX command
is used for the trigger output ports set by the TGP command.
type
=1 to 3 is available for a port only. If you send the command with the same
type
more than once, only the last command is effective.
type
=0 is set for another ports.
If you send the TGP command with
terminal
=1 and
port
=1 to 16, the signal level of
the trigger input terminal is set to physical high.
If you send the TGP command with
terminal
=2, the signal level of the trigger output
terminal is set to logical low.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......