1- 46
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Basics
Programming Tips
To Perform Quasi-Sampling Measurement
The following setup enables to perform a quasi-sampling measurement. Then the
sampling interval will be sum of delay time and step delay time.
•
Sets the sweep measurement mode (MM 2 or MM 16).
•
Sweep start value = Sweep stop value (for WI, WV, or WNX).
•
Sets hold time, delay time, and step delay time (WT).
To Interrupt Command Execution
The E5260/E5270 executes commands in the received order. However, only the
following commands can interrupt the command execution.
Table 1-5
Interrupt Commands
Voltage or current
Measurement trigger (e.g. XE)
Hold
Start value, Stop value
Previous value
: Measurement
Delay
Step delay
Sampling interval = Delay + Step delay
Command
Description
AV
Changes the number of averaging samples during the
measurement.
AIT
Changes the integration time during the measurement.
AB
Aborts the command execution.
*RST
Resets the E5260/E5270 during the command execution.
XE
If the E5260/E5270 has been set to the wait status by the PA or
PAX command, the XE command can be used to release the
wait status. For details, refer to
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......