Keysight E5260/E5270 Programming Guide, Edition 4
3- 69
Programming Examples
Reading Binary Output Data
Reading Binary Output Data
This section provides the example to read binary data. The following program
example:
1. executes high-speed spot measurements
2. reads the measurement data in binary data format
3. rearranges the data and calculates the measured data
4. prints the measured data on the screen
NOTE
Data resolution
The resolution of binary data is as shown below.
•
Measurement data: Measurement range / 50000
•
Output data: Output range / 20000
Note that the resolution of the measurement data is larger than the resolution of the
high resolution A/D converter.
Measurement
Result Example
Id (uA), Status
status = 0
type = 1
mode = 1
channel = 5
sign = 0
range = 0.0001
count = 12010
24.02, 0
Data save completed.
Do you want to perform measurement again?