Keysight E5260/E5270 Programming Guide, Edition 4
1- 31
Programming Basics
Data Output Format
Staircase Sweep,
Multi Channel
Sweep
Block1
[
Block2
] . . . . <terminator>
Block1
is the block of data measured at the first sweep step.
Block2
is the block of
data measured at the second sweep step.
where
Block
consists of the following data:
Data1
[
Data2
] . . . . [
Source_data
]
DataN
(
N
: integer) is the value measured by a channel. The order of
Data
is defined
by the MM command.
Source_data
is the sweep source output value. It is sent if the data output is enabled
by the FMT command.
Pulsed Sweep,
Staircase Sweep
with Pulsed Bias
Block1
[
Block2
] . . . . <terminator>
Block1
is the block of data measured at the first sweep step.
Block2
is the block of
data measured at the second sweep step.
where
Block
consists of the following data:
Data
[
Source_data
]
Data
is the value measured by the channel you specify by using the MM command.
Source_data
is the sweep source output value. It is sent if the data output is enabled
by the FMT command.
Linear Search,
Binary Search
Data_search
[
Data_sense
]<terminator>
This is the data at the measurement point closest to the search target.
Data_search
is the value forced by the search output channel.
Data_sense
is the value measured by the search monitor channel. It is sent if data
output is enabled by the BSVM command for the binary search, or the LSVM
command for the linear search.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......