2- 44
Keysight E5260/E5270 Programming Guide, Edition 4
Remote Mode Functions
Initial Settings
Step delay time
0 s
WT
Trigger delay time
0 s
WT, PT
Trigger mode
XE, TV, TI, or GET
TM
Trigger port
Ext Trig In
Start Measurement
trigger input
TGP
Ext Trig Out
Measurement
Completion trigger
output
TGP
Digital I/O
cleared
TGP
Trigger condition of
Start Step Output Setup trigger
with trigger for first sweep step
TGSI
Type of output trigger
Edge trigger
TGXO, TGSO,
TGMO
Digital I/O port
Output for all port
ERM
Program memory
cleared
a
SCR
Value of internal variable (%In, %Rn)
0
VAR
Data output format
ASCII with header, CR/LF^EOI
FMT
Data output buffer
cleared
BC
Status byte
Only bit 6 is enabled.
*SRE
Error code register
cleared
ERR?
a. Program memory is not cleared by the *RST command or the device clear.
Setup Item
Initial Setting
Commands