Keysight E5260/E5270 Programming Guide, Edition 4
1- 47
Programming Basics
Programming Tips
To Use Programs for Keysight 4142B
Keysight E5260/E5270 supports most of the commands and the data output format
supported by the Keysight 4142B Modular DC Source/Monitor. To reuse the
programs created for the Keysight 4142B, confirm the following and modify the
programs if necessary.
•
To remove all unsupported commands
Some commands are not supported owing to differences in the modules
supported by each instrument. Refer to
that shows the commands not
supported by the E5260/E5270. Do not use these commands.
Perform the linear search or binary search measurement as a substitute for the
analog search measurement that needs the analog feedback unit (AFU).
Use a source/monitor unit (SMU) instead of the voltage source/voltage monitor
unit (VS/VMU). Note that the SMU cannot perform the differential voltage
measurements.
•
FL command
The initial setting of the FL command is different. It is ON for the Keysight
4142B, and OFF for the E5260/E5270.
Add the FL1 command to use the filter.
•
AV command
This command is used to set the A/D converter of the E5260/E5270.
To set the high resolution ADC installed in the E5270B, use the AAD and AIT
commands.
Table 1-6
Modules and Commands Unsupported
Plug-in Module
Commands
Analog Feedback Unit
ASM, AT, ASV, AIV, AVI
High Current Unit
PDM, PDI, PDV
High Voltage Unit
POL
Voltage Source/Voltage Monitor Unit
VM
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......