BSM
The BSM command specifies the search source control mode in the binary search
measurement (MM15), and enables or disables the automatic abort function. The
automatic abort function stops the measurement when one of the following
conditions occurs:
•
Compliance on the measurement channel
•
Compliance on the non-measurement channel
•
Overflow on the AD converter
•
Oscillation on any channel
This command also sets the post search condition for the binary search sources.
After the search measurement is normally completed, the binary search sources
force the value specified by the
post
parameter.
If the search operation is stopped by the automatic abort function, the binary search
sources force the start value after search.
Syntax
BSM
mode
,
abort
[,
post
]
Parameters
mode
:
Source output control mode, 0 (normal mode) or 1 (cautious mode).
If you do not enter this command, the normal mode is set. See
abort
:
Automatic abort function. Integer expression.
1: Disables the function. Initial setting.
2: Enables the function.
post
:
Source output value after the measurement is normally completed.
Integer expression.
1: Start value. Initial setting.
2: Stop value.
3: Output value when the search target value is get.
If you do not specify
post
, the search source forces the start value.
Example
Statements
OUTPUT @E5270;"BSM 1,2,3"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......