4- 80
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
LSV
The LSV command specifies and sets the voltage search source for the linear search
measurement. This command setting is cleared by the LSI command. After search
stops, the search channel forces the value specified by the LSM command.
Syntax
LSV
chnum
,
range
,
start
,
stop
,
step
[,
Icomp
]
Parameters
chnum
:
Search source channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
.
range
:
Output ranging type. Integer expression. The output range will be set to
the minimum range that covers both
start
and
stop
values. Range
changing may cause 0 V output in a moment. For the limited auto
ranging, the instrument never uses the range less than the specified
range. See
start
,
stop
:
Search start or stop voltage (in V). Numeric expression. See
. The
start
and
stop
parameters must have different values.
0 to ±100
0 to ±200 (for E5280B/E5290A)
step
:
Step voltage (in V). Numeric expression.
If
start
<
stop
,
step
must be positive, and if
start
>
stop
,
step
must be
negative. Maximum number of search steps is 1001.
Icomp
:
Current compliance value (in A). Numeric expression. See
. If you do not specify
Icomp
, the previous value is set. Zero
amps (0 A) is not allowed for
Icomp
.
Example
Statements
OUTPUT @E5270;"LSV 1,0,0,20,.5,1E-6"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......