2- 10
Keysight E5260/E5270 Programming Guide, Edition 4
Remote Mode Functions
Measurement Modes
Pulsed Sweep Measurements
Pulsed sweep measurement is performed as shown below. The source channel forces
pulsed sweep voltage or current, and the measurement channel performs one point
measurement at each sweep step.
Figure 2-6
Pulsed Sweep Measurements
1. The pulsed sweep source sets output by the PT, WM, and PWV or PWI
commands. Only one channel can be used for the pulsed sweep source.
2. The pulsed sweep source starts output by a trigger, such as the XE command.
3. After the hold time, the measurement channel starts measurement as shown in
. The channel performs measurement so that the pulse width and pulse
period are kept (the integration time setting is ignored). Only one channel can be
used for measurement.
4. After measurement, the pulsed sweep source forces the pulse base value, and
waits for the rest of the pulse period. Then the pulsed sweep source changes the
output value.
5. The E5260/E5270 repeats 3 and 4 for all sweep steps.
6. After the pulsed sweep measurement, the pulsed sweep source forces the pulse
base value, and keeps it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
Voltage or current
Stop value
Start value
Previous value
Time
: Measurement
Hold time
Pulse period
Trigger (e.g. XE)
PT/WM/PWV/PWI
Pulse base value
Pulse width
Pulse period
Pulse period
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......