4- 38
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
BST
The BST command sets the hold time and delay time for the binary search
measurement.
If you do not enter this command, all parameters are set to 0.
Syntax
BST
hold
,
delay
Parameters
hold
:
Hold time (in seconds) that is the wait time after starting the search
measurement and before starting the delay time for the first search
point. Numeric expression.
0 to 655.35 sec. 0.01 sec resolution.
delay
:
Delay time (in seconds) that is the wait time after starting to force a step
output value and before starting a step measurement. Numeric
expression.
0 to 65.535 sec. 0.0001 sec resolution.
Example
Statements
OUTPUT @E5270;"BST 5,0.1"