Keysight E5260/E5270 Programming Guide, Edition 4
1- 33
Programming Basics
Data Output Format
A
Measurement or source output data type. One bit.
B
Data type. One bit.
C
Measurement or output range. Five bits.
A
Explanation
0
Source output data.
1
Measurement data.
B
Explanation
0
Voltage data.
1
Current data.
C
Explanation
01000 (8)
0.5 V or 1 pA range.
01001 (9)
5 V or 10 pA range.
01010 (10)
100 pA range.
01011 (11)
2 V or 1 nA range.
01100 (12)
20 V or 10 nA range.
01101 (13)
40 V or 100 nA range.
01110 (14)
100 V or 1
μ
A range.
01111 (15)
200 V or 10
μ
A range.
10000 (16)
100
μ
A range.
10001 (17)
1 mA range.
10010 (18)
10 mA range.
10011 (19)
100 mA range.
10100 (20)
1 A range for E5280B/E5290A HPSMU
or 200 mA range for E5291A MPSMU.
11111 (31)
Invalid data is returned.