3- 28
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Examples
Pulsed Sweep Measurements
Pulsed Sweep Measurements
To perform pulsed sweep measurements, use the following commands.
NOTE
Measurement channel performs measurement so that the pulse width and pulse
period are kept. The integration time is automatically set by the instrument, and you
cannot change. For the Keysight E5270B, note that the high-resolution ADC cannot
be used for the pulsed measurements. The AAD/AIT/AV/WT command settings are
ignored.
Function
Command
Parameters
Enables channels
CN
[
chnum
... [,
chnum
] ... ]
Disables channels
CL
[
chnum
... [,
chnum
] ... ]
Sets filter ON/OFF
[FL]
mode
[,
chnum
... [,
chnum
] ... ]
Sets series resistor ON/OFF
[SSR]
chnum
[,
mode
]
Sets pulse timing parameters
PT
hold
,
width
,
period
[,
tdelay
]
Sets auto abort function
[WM]
abort
[,
post
]
Sets pulsed sweep source
PWV
chnum
,
mode
,
range
,
base
,
start
,
stop
,
step
[,
comp
]
PWI
Sets synchronous sweep source
a
[WSV]
chnum
,
range
,
start
,
stop
[,
comp
[,
Pcomp
]]
[WSI]
Forces constant voltage
DV, TDV
chnum
,
range
,
output
[,
comp
[,
polarity
[,
crange
]]]
Forces constant current
DI, TDI
Sets voltage measurement range
[RV]
chnum
,
range
Sets current measurement range
[RI]
chnum
,
range
[RM]
chnum
,
mode
[,
rate
]
Selects measurement mode
MM
4,
chnum
Sets SMU operation mode
[CMM]
chnum
,
mode
Executes measurement
XE
a. The WSV/WSI command must be entered after the PWV/PWI command.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......