4- 96
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
RCV
The RCV command enables the modules that fail the self-test or self-calibration so
that it can receive commands again.
After the RCV command, enter the *OPC? command to confirm that the command
execution is completed.
This command should only be used for servicing the E5260/E5270.
Syntax
RCV [
slotnum
]
Parameters
slotnum
:
Specifies the module to enable. Integer expression. 0 to 9.
0: All failed modules. Default setting.
1 to 8: Module installed in the slot specified by
slotnum
.
9: ADC module installed in the mainframe.
For HPSMU that uses two slots, set the greater slot number. For example, if it is
installed in slot 3 and 4,
slotnum
must be 4.
If
slotnum
specifies the slot that installs no module, this command causes an error.
Example
Statements
OUTPUT @E5270;"RCV 1"
OUTPUT @E5270;"*OPC?"
ENTER @E5270;A
RED
The RED command enables or disables the measurement data display and the setup
data display in the remote mode.
The *RST command or the device clear disables the display.
Syntax
RED
mode
Parameters
mode
:
Data display mode. Integer expression.
0: Disable.
1: Enable.
Example
Statements
OUTPUT @E5270;"RED 1"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......