4- 52
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
DZ
The DZ command stores the settings (V/I output values, V/I output ranges, V/I
compliance values, and series resistor setting) of the specified channels, and sets the
channels to 0 V. The settings can be recovered by using the RZ command. The
stored settings are cleared by using a device clear (HP BASIC CLEAR) command,
*RST, RZ, CL, CA, or *TST?.
Syntax
DZ [
chnum
[,
chnum
...[,
chnum
]...]]
A maximum of eight channels can be set.
Parameters
chnum
:
Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
.
If you specify multiple
chnum
s, the E5260/E5270 sets the channels to 0 V in the
specified order.
If you do not specify
chnum
, the DZ command applies 0 V to all channels where the
output switch is set to ON, in order from higher (output or measurement range) to
lower voltage range.
Remarks
The DZ command sets the specified channels to the following conditions:
Example
Statements
OUTPUT @E5270;"DZ 1,2,3"
Item
Setting
Item
Setting
Output Switch
ON
I Range
See next table
Source Mode
Voltage
Compliance
See next table
Output Voltage
0 V
Filter
Not changed
V Range
Not changed
Series Resistor
Not changed
Previous range
a
a. Range value that was set before the DZ command.
I Range
I Compliance
1 nA to 100
μ
A
same as previous range
range value
over 100
μ
A
100
μ
A
100
μ
A
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......