4- 102
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
SAL
The Keysight E5260 series does not support this command. This command is
available for the Keysight E5270B installed with the high resolution SMU
(HRSMU) and the Atto Sense and Switch Unit (ASU).
Disables or enables the connection status indicator (LED) of the ASU. This
command is effective for the specified channel.
Syntax
SAL
chnum
,
mode
Parameters
chnum
:
Slot number where the HRSMU has been installed. The ASU must be
connected to the HRSMU. Integer expression. 1 to 8.
mode
:
0: Disables the indicator.
1: Enables the indicator. Default setting.
Example
Statements
OUTPUT @E5270;"SAL 1,0"
SAP
The Keysight E5260 series does not support this command. This command is
available for the Keysight E5270B installed with the high resolution SMU
(HRSMU) and the Atto Sense and Switch Unit (ASU). This command is not
effective when the HIGH VOLTAGE indicator of the Keysight E5270B has been
lighted.
Controls the connection path of the ASU. Switches the ASU input resource
(HRSMU or the instrument connected to the AUX input) to be connected to the
ASU output. This command is effective for the specified channel.
After the Keysight E5270B is turned on or the CL command is entered, the ASU
output will be connected to the SMU connector side, but the HRSMU will not be
enabled yet. After this command is entered with
path
=1, the HRSMU specified by
chnum
cannot be used. After this command is entered with
path
=0 or the CN
command is entered, the HRSMU output will appear on the ASU output.Then the
HRSMU output will be 0 V.
Syntax
SAP
chnum
,
path
Parameters
chnum
:
Slot number where the HRSMU has been installed. The ASU must be
connected to the HRSMU. Integer expression. 1 to 8.
path
:
0: The ASU output will be connected to the SMU connector side.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......