XK3118K9 Weighing Indicator Calibration Manual
I. Main Technical Parameters:
Load cell excitation:
DC +5V, can connect 1
~
12 pieces of 350 ohm strain gauge type load cells
Minimum input voltage of each calibration division value: ≥1.5uV/e
Error Distribution coefficient: p
i
=0.5
Load cell connection mode:
6 wires (long wire auto compensation), compensation distance ≤50 meter
Power supply:
AC 110-220V, 50-60Hz,
DC 6V/10Ah free maintenance lead-acid battery, can be charged by the indicator
II. Analog Load Cell Connection:
Specification
Pins
Name
Code
1
Excitation -
E-
2
Feedback -
F-
6
Exci
E+
7
Fe
F+
8
Signal -
S-
9
S+
5
Shield
SHIELD
E+ with F+, E- with F- must be short connected if you use 4 wire shield cable.
III. Calibration
Indicator enters weighing status after self-check. When calibration is needed, open the cover of calibration case on the
back side of indicator, press the calibration touch switch. Then the beep sounds and calibration lamp lights up, which
indicates that the calibration mode can be entered. Then press the calibration button on the front panel. Then the
calibration lamp goes out and enters calibration mode. Detailed calibration steps are as below:
Step
Operation
Display
Remarks
1
Press
【
Calibration
】
【
d *
】
After calibration is finished or interrupted, if need to calibrate again,
the calibration switch should be pressed one more time.
2
Press
【
1
】
【
0
】
Press
【
Input
】
【
d 10
】
Input division, 1/2/5/10/20/50/100 optional,
(factory default setting:
10)
If need compensation calibration, press
【
0
】
press
【
Input
】
enters
to step 7, indicator displays
【
AdloAd1
】
, input weight value of
weights loaded, press
【
Input
】
to confirm
3
Press
【
0
】
Press
【
Input
】
【
dC 0
】
Input decimal bits ( 0
~
4 ), default setting: no decimal 0
4
Press
【
1
】
【
2
】
【
2
】
【
4
】
Press
【
Input
】
【
Pn 01224
】
〖
0 1 2 2 4
Corresponding
VWXYZ
〗
Input system parameter (details reference to the table below)
(factory
default setting: 01224)
V: application (0: non-business scale 1: business scale)
W: stability range
X: zero tracing range
Y
:
manual zero-setting range
Z
:
power on zero-setting range
5
Press
【
3
】
【
0
】
【
0
】
【
0
】
【
0
】
Press
【
Input
】
【
F 30000
】
Input full capacity value
(factory default setting:3000)
If need calibration, input F value
,
then press
【
Input
】
;if press
【
input
】
directly then it enters to step 8
;
if press
【
weight
】
,return back to
weighing status.
(
note
:
full capacity value add 9 division value equals to system preset
overload alarming value.
)
6
Press
【
Input
】
【
noLoAd
】
Zero position confirmation
,
no object on the scale, press
【
Input
】
after
stability symbol light becomes bright
★★★
!!!
When indicator displays
【
noLoAd
】
,
press
【
Weigh
】
button, skip zero calibration, enters to
compensation calibration directly
7
Press
【
2
】
【
0
】
【
0
】
【
0
】
【
0
】
Press
【
Input
】
【
AdloAd1
】
【
20000
】
Add weights, not less than 50% of full capacity, more accurate when it
is near full capacity.
e.g.
:
20000