Press
【
Input
】
to confirm after the stability symbol light becomes bright
8
Press
【
1
】
Press
【
Input
】
【
Adr 01
】
Communication address ( 01
~
26 )
(factory default setting: 1)
9
Press
【
1
】
Press
【
Input
】
【
bt 1
】
Serial communication baud rate ( 0
~
4 ), respectively corresponds to
baud rate: 600/1200/2400/4800/9600 bps
(factory default setting: 1)
10
Press
【
0
】
Press
【
Input
】
【
tF 0
】
Serial communication mode: 0—continuous sending, 1—command
response
(factory default setting: 0)
11
Press
【
2
】
Press
【
Input
】
【
Flt 2
】
Filter strength
(
0
~
6
)
: 0: weakest, 6: strongest
(factory default
setting: 2)
12 Press
【
0
】
【
0
】
【
1
】
【
1
】
【
Y XYZ
】
【
Y 011
】
Special parameter setting
(factory default setting: 0011)
X: 0 Power saving mode OFF 1 Power saving mode ON
X: 0 weight unit is kg; 1 weight unit is ton
Y: 0 data record not use cargo number; 1 data record use cargo
number
Z: 0 data record not use vehicle number; 1 data record use vehicle
number
13
Press
【
Input
】
【
L 0
】
Three section division displays lower limit, the division value will
switch downwards to next part when the gross weight is less than this
value, if calibration division value is 1, it will not switch
(factory
default setting: 0)
14
Press
【
Input
】
【
H030090
】
Three section division displays upper limit, the division value will
switch upwards to the next part when the gross weight is more than this
value, if calibration division value is 100, it will not switch
(factory
default setting: division value switch not valid)
15
【
20000
】
Calibration finishes, return back to weighing status
Note 1: in step 6 and step 7, if press
【
Weigh
】
, it means this step does not work, it enters to the next step directly
Note 2: in the calibration process, press
【
Weigh
】
in other steps, it will return back to weighing status
Note 3: Pn parameter setting ways
IV.
Operation of Parameters Calibrated
V. Timing shut-down and random unlock
Press
【
PRINT SET
】
under weighing status. Input password
【
88
】
. Then press
【
INPUT
】
. The indicator displays
【
SECU 0
】
, which means that enters timing shut-down setting menu (the password and date of timing shut-down can
be changed).
The password of timing shut-down should be kept carefully. The factory setting password is
Input value
0
1
2
3
4
5
6
7
8
9
Unit
W: Stability range
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
uV
X: zero tracing range
0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
E
Y:
manual
zero-setting
range
0
2%
4%
10%
20%
No
limit
—— —— —— ——
F.S
Z: Start-up zero-setting
range
0
2%
4%
10%
20%
No
limit
—— —— —— ——
F.S
No.
Function
Operation
1
Inquire the recent
date of calibration
Press
【
SET PRINT
】
, input
【
9
】
【
0
】
→ Press
【
Input
】
to display“d**.**.**”(Date)→Press
【
Input
】
to display “t**.**.**”(Time)→Press
【
Input
】
to return to weighing status.
2
Backup the
calibration
parameters
Press the touch button (the beep sounds and calibration lamp fashes) → Press
【
SET
】
and input
【
9
】
【
1
】
→ Press
【
Input
】
to display “blbc 0”→ Input
【
1
】
(any number
except zero)→ Press
【
Input
】
to backup the calbiration parameters, and return to weighing
status.
3
Recover the
calibration
parameters
Press touch calibration switch (the beep sounds and calibration lamp fashes)→Press
【
SET
】
and enter
【
9
】
【
2
】
→Press
【
Input
】
to display“blHF 0”→Enter
【
1
】
( any number
except zero )→Press
【
Input
】
to recover the calibration parameters, and return to weighing
status. The recovery of the calibration parameters is deemed as one complete calibration.
4
Inquire the date to
backup the
calibration
parameters
Press
【
SET PRINT
】
and input
【
9
】【
3
】
→Press
【
Input
】
to
display“d**.**.**”(Date)→Press
【
Input
】
to display “t**.**.**”(Time)→Press
【
Input
】
to return to weighing status.