![background image](http://html.mh-extra.com/html/keithley/6430/6430_instruction-manual_661765494.webp)
B-6
Status and Error Messages
Number
Error message
Event
1
Status register
2
Bit
+305
+306
+310
Waiting in trigger layer
Waiting in arm layer
Entering idle layer
SE
SE
SE
Operation Event
Operation Event
Operation Event
5
6
10
+408
+414
Questionable events:
Questionable Calibration
Command Warning
SE
SE
Questionable Event
Questionable Event
8
14
+500
+501
+502
+503
+504
+505
+506
+507
+508
+509
+510
Calibration errors:
Date of calibration not set
Next date of calibration not set
Calibration data invalid
DAC calibration over
fl
ow
DAC calibration under
fl
ow
Source offset data invalid
Source gain data invalid
Measurement offset data invalid
Measurement gain data invalid
Not permitted with cal locked
Not permitted with cal un-locked
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
EE
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
3
3
3
3
3
3
3
3
3
3
3
+601
+602
+603
+604
+605
+606
Lost data errors:
Reading buffer data lost
GPIB address lost
Power-on state lost
DC calibration data lost
Calibration dates lost
GPIB communication language lost
EE
EE
EE
EE
EE
EE
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
3
3
3
3
3
3
+700
+701
+702
Communication errors:
Invalid system communication
ASCII only with RS-232
Preamp Timeout
EE
EE
EE
Standard Event
Standard Event
Standard Event
3
3
3
1
EE = Error Event
SE = Status Event
SYS = System Error Event
2
Use following queries to read status registers:
Standard Event: *ESR?
Operation Event: STAT:OPER?
Measurement Event: STAT:MEAS?
Questionable Event: STAT:QUES?
Table B-1 (cont.)
Status and error messages
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......