Cases V and VI: Measure one function, Output Auto-Off disabled, and no
source setting changes.
Auto-Zero:
Enabled
Trigger Latency:
500µsec max
Auto Zero:
Disabled
Trigger Latency:
100µsec max
The source turns on as soon as the output is turned on and remains on until the source is
turned off. As shown in the two timing diagrams, the static source remains on for every mea-
surement cycle. The Source-Delay portion of the SDM cycle is omitted. With Trigger Delay set
to zero, Trigger Latency is the time from when the trigger event occurs to when the Source-
Meter begins an A/D conversion.
Trigger
Latency
Trigger
Delay
Trigger
Event
Source On
A/D
Conversion
(signal phase)
A/D
Conversion
(ref phase)
A/D
Conversion
(ref zero phase)
F
i
g
u
r
e A
-5
Case V timing
diagram
Trigger
Latency
Trigger
Event
Source On
Trigger
Delay
A/D
Conversion
(signal phase)
F
i
g
u
r
e A
-
6
Case VI timing
diagram
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......