Basic Source-Measure Operation
Remote command measure only
Table 3-8 summarizes the basic command sequence for measure only. The steps outlined
correspond to those in the
Front panel measure only
sequence above. These commands set up
the SourceMeter for measure only voltage measurements up to 20V as follows:
•
Measure function: volts
•
Source function: current
•
Source mode: fixed
•
Source range: minimum
•
Source value: 0mA
•
Measure range: 20V
•
Compliance 25V
T
a
b
le 3
-8
Measure only programming example
Step
1
Action
Commands
2,3
Comments
*RST
Restore GPIB defaults.
1
Select measure, source functions.
:SOUR:FUNC CURR
Current source function.
:SOUR:CURR:MODE FIXED
Fixed current source mode.
:SENS:FUNC “VOLT”
Volts measure function.
2
Set source and compliance.
:SOUR:CURR:RANG MIN
Lowest source range.
:SOUR:CURR:LEV 0
0
µ
A source level.
:SENS:VOLT:PROT 25
25V compliance.
3
Select volts measure range.
:SENS:VOLT:RANG 20
20V range.
5
Turn on output.
:OUTP ON
Output on before measuring.
6
Read data.
:READ?
Trigger, acquire reading.
7
Turn off output.
:OUTP OFF
Output off after measuring.
1
Steps correspond to front panel steps listed previously in
Front panel measure only
. DUT should be connected to SourceMeter before
running program.
2
Commands must be sent in order given.
3
Instrument must be addressed to talk after :READ? to acquire data.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......