6-2
Range, Digits, Speed, and Filters
Range and digits
Range
The selected measurement range affects the accuracy of the measurements as well as the
maximum signal that can be measured. Note that with the output off, dashed lines are displayed
(i.e., --.---- µA), to indicate that measurements are not being performed.
Available ranges
The SourceMeter consists of a mainframe and a Remote PreAmp. The mainframe can be
used with or without the Remote PreAmp. However, when not using the Remote PreAmp, the
lower current ranges and higher resistance ranges are not available. Table 6-1 lists the available
ranges for the SourceMeter.
Table 6-1
Model 6430 ranges
Voltage Ranges
Current Ranges
Ohms Ranges
With Remote
PreAmp
Mainframe
Only
With Remote
PreAmp
Mainframe
Only
With Remote
PreAmp
Mainframe
Only
200V
20V
2V
200mV
200V
20V
2V
200mV
100mA
10mA
1mA
100µA
10µA
1µA
100nA
10nA
1nA
100pA
10pA
1pA
100mA
10mA
1mA
100µA
10µA
1µA
<2
Ω
*
20
Ω
200
Ω
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
200M
Ω
2G
Ω
20G
Ω
200G
Ω
2T
Ω
20T
Ω
<2
Ω
*
20
Ω
200
Ω
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
*Manual ohms mode only
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......