5-16
Source-Measure Concepts
Where within the boundaries the SourceMeter operates depends on the load (DUT) that is
connected to the output. Figure 5-8 shows operation examples for resistive loads that are 2k
Ω
and 800
Ω
, respectively. For these examples, the SourceMeter is programmed to source 10V
and limit 10mA.
In Figure 5-8A, the SourceMeter is sourcing 10V to the 2k
Ω
load and subsequently mea-
sures 5mA. As shown, the load line for 2k
Ω
intersects the 10V voltage source line at 5mA.
Figure 5-8B shows what happens if the resistance of the load is decreased to 800
Ω
. The
DUT load line for 800k
Ω
intersects the current compliance limit line placing the SourceMeter
in compliance. In compliance, the SourceMeter will not be able to source its programmed volt-
age (10V). For the 800k
Ω
DUT, the SourceMeter will only output 8V (at the 10mA limit).
Notice that as resistance decreases, the slope of the DUT load line increases. As resistance
approaches infinity (open output), the SourceMeter will source virtually 10V at 0mA. Con-
versely, as resistance increases, the slope of the DUT load line decreases. At zero resistance
(shorted output), the SourceMeter will source virtually 0V at 10mA.
Regardless of the load, current will never exceed the programmed compliance of 10mA.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......