Calibration
19-21
5.
When the instrument is
fi
nished performing offset calibration for all ranges, it will
return to the calibration menu.
Step 5: Lock out calibration
1.
From the calibration menu, select LOCK, then press ENTER. The instrument will
return to the main menu.
2.
Press EXIT to return to normal display.
Changing the password
Follow the steps below to change the password:
1.
Press the MENU key, then choose CAL and press ENTER. The instrument will display
the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
2.
Select UNLOCK, then enter the password. (Default: 006430.)
3.
Select CHANGE-PASSWORD, and then press ENTER. The instrument will display
the following:
New Pwd: 006430
Use
,
▲
,
▼
,
, ENTER or EXIT.
4.
Using the range keys, and the left and right arrow keys, enter the new password on the
display.
5.
Once the desired password is displayed, press the ENTER key to store the new
password.
Resetting the calibration password
If you lose the calibration password, you can unlock calibration by shorting together the
CAL pads, which are located on the display board. Doing so will also reset the password to the
factory default (006430).
▲
▲
▲
▲
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......