17-52
SCPI Command Reference
OUTPut subsystem
This subsystem is used to control the output of the selected source, and the interlock. These
commands are summarized in Table 17-4.
Turn source on or off
[:STATe] <b>
:OUTPut[1][:STATe] <b>
Turn source on or off
Parameters
<b> =
0 or OFF
Turn source off (standby)
1 or ON
Turn source on (operate)
Query
:OUTPut?
Query state of source
Description
This command is used to turn the source output on or off. Measurements
cannot be made while the source is off.
Turning the source off places the SourceMeter in the idle state. The only
exception to this is when source auto clear is enabled. In this mode, the
source turns on during each source phase of the SDM cycle and then turns
off after each measurement. See :SOURce[1]:CLEar:AUTO in
SOURce
subsystem
.
NOTE
The :SOURce:CLEar command will also turn the source off.
Interlock control
:STATe <b>
:OUTPut[1]:INTerlock:STATe <b>
Control hardware interlock
Parameters
<b> =
0 or OFF
Disable interlock
1 or ON
Enable interlock
Query
:STATe?
Query state of interlock
Description
This command is used to enable or disable the hardware interlock. When
enabled, the source cannot be turned on unless the interlock line (pin 8 of
the rear panel Interlock - Digital I/O connector) is pulled to a logic low
state. When the interlock line goes to a logic high state, the source turns off.
See Section 12,
Digital I/O port
and
Safety interlock
for details using inter-
lock with a test
fi
xture.
When disabled, the logic level on the interlock line has no effect on the out-
put state of the source.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......