Getting Started
1-9
Mainframe rear panel summary
The rear panel of the Model 6430 is shown in Figure 1-2.
Source-measure connectors:
CAUTION
When using the Remote PreAmp, DO NOT use the INPUT/OUTPUT HI
and 4-WIRE SENSE HI banana jacks. Using these source-measure
banana jacks will result in errors and/or noise, and possible damage to the
Remote PreAmp. Use the input/output HI and sense HI terminals on the
Remote PreAmp.
INPUT/OUTPUT HI and LO
Use to source-measure volts, amps, and ohms.
4-WIRE SENSE HI and LO
Use for 4-wire remote sensing.
V,
Ω
GUARD
Driven guard for guarded measurements.
GUARD SENSE
Use to correct for IR drops in Guard Output lead.
Earth (chassis) ground screw.
WARNING
INPUT/OUTPUT LO is not internally connected to the chassis and can
float up to 42V peak above chassis ground.
The maximum allowable voltage differentials between the various terminals are provided on
the rear panel of the mainframe (see Figure 1-2):
Low voltage differentials
— The maximum low voltage differentials are 5V peak. How-
ever, to achieve rated accuracy, do not exceed 1V peak on the 100µA through 100mA ranges.
For the lower current ranges (1pA through 10µA), do not exceed the maximum allowable resis-
tance. The maximum resistance for each of these current ranges are provided in Section 2 (see
Connections to DUT, 4-wire sensing specification clarifications)
.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
INTERLOCK-
DIGITAL I/O
Figure 1-2
Model 6430
rear panel
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......