F-2
Measurement Considerations
NOTE
This appendix summarizes considerations to make safe, accurate measurements
using the Model 6430. For comprehensive information on these and additional mea-
surement considerations, refer to the Low Level Measurements handbook, which is
available from Keithley.
Floating measurement safety concerns
The Model 6430 can perform
fl
oating measurements up to 42V above chassis ground. Even
though 42V is not considered a lethal level, it is high enough to cause a shock.
Figure F-1 shows two examples where the Model 6430
fl
oats at 40V above chassis ground.
Keep in mind the outer shells of the triax connectors on the Remote PreAmp are connected to
input/output LO. Therefore, 40V is present on those shells as well as the outer casings of the
triax cable(s). If meter LO is connected to a noise shield, then 40V will also be present on that
shield. Typically, a test circuit is enclosed in a test
fi
xture that is connected to a safety earth
ground. The test connection drawings in Section 2 show how a test
fi
xture should be used.
WARNING
To prevent a shock hazard and possible damage to the Model 6430, never
exceed 42V peak between input/output LO and chassis ground.
WARNING
To prevent injury from electric shock, DO NOT touch the triax cables (or
connectors) of the Remote PreAmp while the SourceMeter is turned on or
any external source is turned on.
80V
(R
1
= R
2
)
R
1
HI
40V
V
LO
+
-
R
2
6430
Voltmeter
40V
R
1
40V
A
+
-
R
2
6430
Ammeter
R
3
A. Voltage measurement
B. Current measurement
Figure F-1
Floating measurements
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......